First oscilloscope lab: probe compensation, DC vs AC coupling and reading Vpp

Principles and common slips of the function generator and oscilloscope lab, using the values the scope displayed.

The first electronics lab moves on from the bench supply and multimeter to the function generator and the oscilloscope. The supply and multimeter part is covered with real data in the ammeter guide; this is the second half. The numbers in the tables are copied exactly as the scope's measurement panel showed them, and the waveform plots are redrawn from the generator settings.

In this guide 7

1. What the lab is for

  • Compensate a 10x probe and see what an uncompensated probe does to a waveform.
  • Read Vpp and frequency of sine, square and triangle waves off the screen.
  • Add a DC offset and see the difference between DC and AC coupling.
  • Find out which controls leave the readings alone (vertical position, time base) and which ones get in the way.

2. How it works

Probe compensation. A 10x probe divides the signal by ten with 9 MΩ in the probe and the scope's 1 MΩ input. The scope input and the cable also have a capacitance C2, so with resistors alone the higher-frequency parts of a signal would be cut more. An adjustable C1 across the 9 MΩ makes the two time constants equal.

R1 9 MΩ C1 cable R21 MΩ C2 Vin probe scope input
A 10x probe as a circuit. The arrow marks C1, the trimmer you turn; C2 is the scope input plus the cable capacitance.
R1C1 = R2C2compensation: equal time constants divide by ten at every frequency
(Vscope)/(Vin) = (R2)/(R1 + R2) = (1)/(10)attenuation once compensated

Too much C1 makes the corners of a square wave spike and settle (over-compensated); too little rounds them off (under-compensated). With example values the over-compensated corner jumps to 1.23 V, the under-compensated one to 0.74 V, and both settle to 1 V with a time constant of about 100 µs.

A 0–1 V square wave seen through a 10x probe (calculated)

Example values: R1 9 MΩ, R2 1 MΩ, C2 100 pF. C1 = 14 pF (over), 11.1 pF (compensated), 8 pF (under).

The report's before-compensation screen showed exactly that spike-and-settle shape at every edge, and the corrected screen a flat square wave. A sine keeps its shape and only changes slightly in amplitude, so always check compensation on a square wave.

DC and AC coupling. DC coupling shows the signal as it is. AC coupling puts a series capacitor in the input to remove the mean (the DC part). That capacitor and the input resistance form a high-pass filter, so only signals well above its corner pass unchanged. The corner varies by model but is usually a few hertz to about 10 Hz, which hardly matters at 1 kHz.

fL = (1)/(2πRinCAC)the high-pass corner made by the AC coupling capacitor
Vhigh = Vpp(1 − D), Vlow = −Vpp·Dlevels of a duty-D square wave after AC coupling

A sine's mean equals its offset, so on AC coupling it drops to sit around 0 V. A 30% duty square wave's mean is not its midpoint but 1.2 V, so on AC coupling its levels become lopsided.

CouplingHigh levelLow levelMean
DC4.0 V0.0 V1.2 V
AC2.8 V−1.2 V0 V
Sine, 4 Vpp, 2 V offset
Square, 4 Vpp, 30% duty, 2 V offset

Try it: the lab's scope screen

The lab scope's 14-division screen. Change the generator, the probe, the coupling and the scales; the measurement panel follows. It starts on the 30% duty square wave capture (4 Vpp, 2 V offset).

Vpp
—
Max
—
Min
—
Mean
—
Frequency
—
Periods on screen
—

Things to try:

  • Switch to AC coupling: the 30% square wave loses its 1.2 V mean and its levels become 2.8 V and −1.2 V, lopsided. Vpp does not change.
  • Set the time base to 100 µs/div: only 1.4 periods fit and the frequency turns to "?", as it did in the lab.
  • Pick an over- or under-compensated probe and the corners change the way they did before the trimmer was turned.
  • Shrink V/div or move the trace until it runs off the screen: Vpp then reports the clipped height.
  • On AC coupling, take the frequency below a few tens of hertz and the flat tops droop through the coupling capacitor (its corner is assumed to be 5 Hz here).

The probe curves use the example values above (R1 9 MΩ, R2 1 MΩ, C2 100 pF). For a sine's rms and dBV, see the oscilloscope widget on the dB calculator.

3. The setup

Function generator Vs CH1CH2 ×10×10 CALGND probe common ground (probe clips)
The setup. With both channels on the same point you can compare them directly. Compensate a probe by clipping its tip to the CAL terminal.

Probe the generator output (Vs) and clip every ground lead to the same ground. Putting the same signal on both channels lets you compare channels and probes, which is how a misbehaving channel shows up, as below.

4. Step by step

  1. Set the probe switch to 10x and the scope channel's probe setting to 10X as well. If they differ, every reading is off by ten.
  2. Clip the tip to the CAL terminal (usually a 1 kHz square wave) and the ground clip next to it, then turn the trimmer until the corners are flat. Do each channel.
  3. Set the generator to a 1 kHz, 4 Vpp sine with 0 V offset, and check its output load setting is High-Z, not 50 Ω.
  4. Choose a time base that shows at least two periods and a vertical scale that fills at least half the screen. Keep the trigger level between the waveform's minimum and maximum.
  5. Turn on the Vpp and frequency measurements, note them, then add a 2 V offset and switch between DC and AC coupling.
  6. Repeat with square waves (50% and 30% duty) and a triangle. Also try moving the vertical position and changing the time base, and note what the readings do.

5. The readings

Values the scope showed for the 4 Vpp sine. In the setting column 0 V and 2 V are the offset, DC and AC the coupling, "Moved" the capture with the vertical position moved, and 100 µs/div the one with the time base changed. Everything is at 1 kHz except the 10 kHz row, and the ratio is CH2 ÷ CH1. The time base was 500 µs/div at 1 kHz and 50 µs/div at 10 kHz. One photo was too blurred to read Vpp, so it is left out (—).

Amplitude

SettingCH1 VppCH2 VppRatio
0 V4.015 V2.975 V0.741
2 V, DC3.995 V2.985 V0.747
2 V, AC4.020 V2.990 V0.744
Moved4.005 V3.125 V0.780
100 µs/div———
10 kHz4.080 V3.435 V0.842

Frequency

SettingCH1 FCH2 F
0 V1.003 kHz1.000 kHz
2 V, DC1.000 kHz1.000 kHz
2 V, AC1.000 kHz1.000 kHz
Moved1.000 kHz1.000 kHz
100 µs/div??
10 kHz10.00 kHz10.00 kHz
Sine Vpp as displayed (line: the 4 Vpp generator setting)

Dots: CH1 and CH2 readings. The four dots at 1 kHz are different settings at the same frequency.

The square and triangle waves were measured on CH1 only. 30% is the square wave's duty cycle and the triangle's symmetry (the share of the period spent rising); only the 30% square wave had a 2 V offset, the others 0 V. Every one of these captures showed 1.000 kHz; "vs setting" is the difference from the 4 Vpp setting.

WaveformCH1 Vppvs setting
Square 30%, DC4.186 V+0.186 V
Square 30%, AC4.210 V+0.210 V
Square 50%4.240 V+0.240 V
Triangle 30%4.010 V+0.010 V

6. Reading the results

  • Coupling does not change Vpp: with the 2 V offset sine, CH1 read 3.995 V on DC coupling and 4.020 V on AC coupling. Coupling only moves the trace (its mean). The four CH1 readings at 1 kHz span 3.995 to 4.020 V, within 1% of the 4 Vpp setting.
  • Only CH2 read low: CH2 was 0.74 to 0.78 times CH1 at 1 kHz and 0.84 times at 10 kHz. A wrong attenuation setting is off by a factor of ten and does not change with frequency, so it cannot explain this alone. A ratio that moves with frequency is typical of a probe that is not fully compensated; the report also does not say exactly where CH2 was connected, so the connection point and the probe contact are candidates too. Clip both probes to the same point, then swap them, to tell whether the probe or the channel is to blame.
  • Square waves read above 4 V: 4.186 to 4.240 V for the square waves against 4.010 V for the triangle. Vpp is maximum minus minimum, so any overshoot, ringing or single noise spike at an edge goes straight into it. Probe compensation and a long ground lead shape the edges, so suspect those first. For the height of the flat parts, use an amplitude (top − base) measurement or cursors.
  • When frequency shows "?": at 100 µs/div both channels showed "?" for frequency. With 14 divisions across, the screen held only 1.4 periods, too few to time one. At 500 µs/div it showed 7. Show at least two periods to measure frequency.
  • Vertical position does not affect the readings: it only moves where the trace is drawn. If the trace runs off the top or bottom, though, Vpp reports the clipped height, so keep the whole waveform on screen.
  • Small frequency differences: 1.003 kHz against 1.000 kHz is the resolution of the scope's frequency measurement. When comparing with the generator setting, only the displayed digits mean anything.

7. Common slips

  • The probe switch (1x or 10x) and the channel's probe setting disagree, so readings are off by ten.
  • The generator's output load is set to 50 Ω while it drives a High-Z scope input, so the screen shows twice the set amplitude; a 50 Ω load on a High-Z setting halves it.
  • The trigger level is outside the waveform, so the picture rolls. Check the trigger level again after adding an offset or changing coupling.
  • Ground clips go to different potentials. Scope ground is normally tied to mains earth, so clipping it to another point in the circuit shorts that point to ground.
  • On AC coupling, a very low-frequency square wave sags across its top, and that gets mistaken for a signal problem. Near the corner frequency the coupling capacitor discharges and the waveform droops.

Convert Vpp to rms or dBV with the dB calculator, work out RC time constants for AC coupling and probe compensation with the RC filter calculator, and check a multimeter's accuracy on the same signal with the multimeter accuracy calculator.